How many PCR reactions are needed per patient?
PCR Setup / PCR reactions per sample
Every
sample is applied to 3 (three) master mixes: MA
Bac for 16S, MA Yeasts for 18S,
and MA Control for the extraction
control.
All master mixes need at least one negative control and MA Bac and MA Yeasts have two PCR positive controls: P1 to verify the function
of the PCR and P2, a dilution of P1, as an indicator for the sensitivity
reached.
For MA Control there is no
positive control required.
Example
for the calculation of the required PCR reactions for the simultaneous analysis
of four samples below:
| MA Bac
(16S) | MA Yeasts
(18S) | MA Control |
Sample(s) | 4 | 4 | 4 |
Negative control (DNA-free water) | 1 | 1 | 1 |
Positive control P1 | 1 | 1 | 0 |
Positive control P2 | 1 | 1 | 0 |
Total number of reactions | 7 | 7 | 5 |
PCR
Setup Guide (click on attachment)
Related Articles
How to interpret the PCR results?
PCR results Interpretation Amplicons can either be analyzed by conventional gel electrophoresis with bands displaying the size of the amplicons (460bp for the MA Bac and 310bp for the MA Yeasts). Alternatively, a real-time PCR cycler can be used, and ...
How many sequencing reactions are needed per sample?
Sequencing For every sample with a positive 16S PCR, there are two sequencing reactions based on two sequencing primers, SeqGN16 and SeqGP16. This allows the identification of up to two pathogens, mainly separated into Gram-negative and Gram-positive ...
Does a small number of samples per extraction (manual or automated in the SelectNA™plus instrument) reduce the total amount of reactions per kit?
Number of reaction / extraction The number of reactions indicated in the kit size refers to the number of extractions e.g. a kit with 48 reactions = 48 extractions. The limitation arises concerning the reactions in the master mixes: Since every PCR ...
Positive Control DNA (P1) Datasheet
The latest versions of the following datasheets are available for download below: Positive Control DNA (P1) Datasheet (English)
Which real-time PCR cyclers are validated for SepsiTest™-UMD and Micro-Dx™?
PCR Cyclers The following cyclers are validated: LightCycler® 1.5, 2.0, 96, 480 and Nano, Roche DNA Engine Opticon®, CFX96™, BioRad Mx3000P®, Mx3005P®, Stratagene ABI 7500 Fast®, Life Technologies Rotor-Gene®, Qiagen peqStar 96Q, peqlab Not suitable ...